  keysight technologies infiniium 90000 series oscilloscopes data sheet
02 | keysight | infiniium 90000 series oscilloscopes - data sheet table of contents why choose keysight oscilloscopes for your toughest high-speed measurement challenges? ?????????????????????????????????????????????????????????????????????????????????????????????????????????????? 03 engineered for unmatched real-time measurement accuracy ???????????????????????????????????????????????????????????????? 04 keysight infiniium portfolio ?????????????????????????????????????????????????????????????????????????????????????????????????????????????????????????? 19 infiniium 90000a series ?????????????????????????????????????????????????????????????????????????????????????????????????????????????????????????????? 20 infiniimax ii: the worlds best high-speed probing system just keeps getting better ?????????????????????????? 22 infiniium 90000a series oscilloscopes ??????????????????????????????????????????????????????????????????????????????????????????????????????? 24 infiniimax ii series ???????????????????????????????????????????????????????????????????????????????????????????????????????????????????????????????????????? 34 infiniium 90000 series ordering information ?????????????????????????????????????????????????????????????????????????????????????????????? 36 keysight oscilloscopes ?????????????????????????????????????????????????????????????????????????????????? ?????????????????????????????????????????????? 39 evolving since 1939 ????????????????????????????????????????????????????????????????????????????????????????????????????????????????????????????????????? 41
03 | keysight | infiniium 90000 series oscilloscopes - data sheet why choose keysight oscilloscopes for your toughest high-speed measurement challenges? as an engineer youre no stranger to tough challenges that help you exceed your customers needs and expectations ? deploying your next design successfully is even more difficult with todays high speed technologies ? data signal eye openings become smaller and measurement error from your oscilloscope becomes less tolerable ? keysight technologies, inc ? is committed to providing the best measurement solutions for these critical challenges ? the keysight infiniium 90000 series oscilloscopes are engineered to give you unmatched real-time measurement accuracy so you can: 1 ? use your jitter budget in your design, not on your scilloscope ? 2 ? pass todays demanding compliance tests more quickly ? 3 ? debug your toughest designs with confidence ? how much time span can i capture? sampling rate 20 mpts of memory 50 mpts of memory 100 mpts of memory 200 mpts of memory 500 mpts of memory 1 gpts of memory 40 gsa/s 500 s 1.25 ms 2.5 ms 5.0 ms 12.5 ms 25.0 ms 20 gsa/s 1 ms 2.5 ms 5.0 ms 10.0 ms 25.0 ms 50.0 ms note: time span capture = memory depth x 1/ sampling rate ? 90000a series infiniium oscilloscopes model real-time bandwidth on 4 ch maximum sampling rate on 4 ch standard memory maximum memory noise floor at 50 mv/div 91304a 13 ghz 40 gsa/s on 4 ch 20 mpts on 4 ch 1 gpts on 4 ch 1.73 mvrms 91204a 12 ghz 40 gsa/s on 4 ch 20 mpts on 4 ch 1 gpts on 4 ch 1.45 mvrms 90804a 8 ghz 40 gsa/s on 4 ch 20 mpts on 4 ch 1 gpts on 4 ch 1.15 mvrms 90604a 6 ghz 20 gsa/s on 4 ch 1 20 mpts on 4 ch 1 gpts on 4 ch 0.98 mvrms 90404a 4 ghz 20 gsa/s on 4 ch 1 20 mpts on 4 ch 1 gpts on 4 ch 0.79 mvrms 90254a 2.5 ghz 20 gsa/s on 4 ch 1 20 mpts on 4 ch 1 gpts on 4 ch 0.64 mvrms 1 ? dsa model numbers come standard with 50 mpts of memory on 4 ch ? engineered for unmatched real-time measurement accuracy
04 | keysight | infiniium 90000 series oscilloscopes - data sheet the industrys lowest noise floor leveraging the company expertise in rf design, keysight has invested in key technology blocks like our proprietary faraday caged front end to significantly reduce our inherent scope noise floor ? the result: the lowest noise floor available on any real-time oscilloscope from 2 ? 5 g hz to 12 ghz ? the industrys deepest memory with 1 gbyte of memory, low-frequency jitter components can be more quickly resolved in a single measurement ? statistical accuracy is improved with more data collection ? keysights integrated deep memory remains responsive and allows more comprehensive testing, supporting pattern lengths up to prbs23 for accurate transmitter and receiver results ? we add full bandwidth probing and accurate de-embedding and equalization software the performance of keysights oscilloscopes is matched by the superiority of our probing, de-embedding and equalization offerings ? maintain full bandwidth performance to the probe tip with our infiniimax probing solutions ? r ender waveforms anywhere in the digital serial link with our hardware accelerated n5465a infiniisim waveform transformation toolset ? configurable system modeling allows you to remove the deleterious effects of unwanted channel elements, simulate waveforms with channel models inserted, view waveforms in physically unprobeable locations, and compensate for loading of probes and fixtures ? the n5461a serial data equalization software allows you to model equalization techniques in real time ? use your jitter budget in your design, not on your oscilloscope keysights infiniium 90000 series oscilloscopes offer the industrys lowest noise floor and most accurate real-time jitter measurements available on scopes of this bandwidth class ? complete with full-bandwidth probing solutions and hardwareaccelerated de-embedding and equalization techniques, keysight oscilloscopes are the best oscilloscope solution for todays demanding high-speed measurements ? engineered for unmatched real-time measurement accuracy
05 | keysight | infiniium 90000 series oscilloscopes - data sheet pass todays demanding compliance tests more quickly offering the industrys widest range of available compliance applications to provide fast setup for complete, automated compliance and margin testing and reporting, the keysight 90000 series scopes have become the go-to tool for test houses worldwide ? our experts serve on the industry standards committees, and our oscilloscopes are certified on todays high speed serial data standards ? plus our 1g memory supports real-time testing to pattern lengths of prbs23 to stress your design to the max ? choose from a wide range of complete compliance applications choose from the industrys widest range of complete applications for the infiniium 90000 series to ensure compliance to the leading industry standards, including sata, pci express ? , ethernet, usb, and more ? comprehensive set-up wizards and full automation of the required testing take the guesswork out of demonstrating compliance quickly ? get further insight with our protocol and analysis decode available on pci express, sata and usb ? put keysights experts on your team keysights measurement experts sit on the industry standards committees and help define the compliance requirements ? they make sure our tools deliver exactly to the standards ? you get the benefit of years of training and experience on every measurement you make ? free up valuable engineering resources set-up wizards combined with intelligent test filtering make it simple to ensure the right tests are being run ? comprehensive html reports with visual documentation and pass/ fail results guarantee that critical information is retained on each test ? technicians can run complete and accurate testing on their own, freeing valuable engineering resources ? precisionprobe keysights n2809a-1nl precisionprobe software quickly characterizes and compensates the frequency response of any path to the 90000 series input ? precisionprobes patented technology uses the < 15ps edge from the 90000 series oscilloscope to: C measure input impedance and response of any probe and the loss of any cablequickly correct from probe and cable loss(without extra instruments such as vna or tdr) C quickly correct from probe and cable loss(without extra instruments such as vna or tdr) C correct probing issues such as phase nonlinearity, magnitude non-flatness, and see the effect of probe loading C quickly gain insight into impedance/capacitance that defines your connection engineered for unmatched real-time measurement accuracy (continued) precisionprobe uses keysights proprietar 200 ghz indium phosphide process to create a fast edge for characterization with precisionprobe ?
06 | keysight | infiniium 90000 series oscilloscopes - data sheet debug your toughest designs with confidence the 90000 series boasts an ever-expanding set of measurement applications for serial debugging and protocol viewing, jitter testing, advanced triggering, measurement customization, and rapid automation ? put the power of the scope to work for your unique debug and analysis challenges ? streamline your debug and analysis tasks with the industrys widest range of application software whether you need to trigger and decode serial buses, iron out the kinks in your memory designs, or see fft based spectrum analysis of your signal, the infiniium 90000 series has application software to help ? our serial protocol views are unique to oscilloscopes, and our ddr debug tools support multiple generations of the standard ? quickly access additional features from the scopes standard menus ? customize your scope for even more efficiency the n5414b-1nl infiniiscan event identification software makes unique capabilities like zone qualify and generic serial triggering possible ? rapidly automate any scope measurement using the n5467b-1nl user defined application and have it appear seamlessly in your scopes menu ? customize your infiniium further by taking full advantage of myinfiniium (standard on all 90000a series oscilloscopes) ? use myinfiniium to deliver automated measurements, execute customized scripts, save screenshots, or load your favorite setup ? add measurement capability with matlab compatibility if we havent provided exactly what you need, take customization to a new level with matlab (option 062) - a data analysis software environment and scripting language with over 1,000,000 users today ? use matlab to design and apply your own filters to oscilloscope signals, graphically visualize oscilloscope signals in 2-d and 3-d plots, automate measurements, and build test applications ? add the n5430a-1nl user defined function software to your scope to seamlessly integrate your custom functionality into the infiniium 90000 menus so results are displayed on the scope screen ? keysight is the only t&m manufacturer today that sells and supports matlab as its own product as part of a complete t&m solution ? engineered for unmatched real-time measurement accuracy (continued)
07 | keysight | infiniium 90000 series oscilloscopes - data sheet trigger and view on-screen serial decode of i 2 c packets frequency domain analysis infiniiums built-in fft allows users to quickly and easily analyze the frequency components of their signals ? both fft magnitude and phase can be displayed and can be combined with other built-in math functions or matlab based measurements ? a resolution bandwidth of 6 khz is supported with the standard 10 megabytes of acquisition memory at the maximum sample rate of 40 gsa/s ? with optional acquisition memory installed resolution bandwidths of 2 khz can be obtained ? standard windowing of hanning, flattop and rectangular are supported along with cursor based power measurements ? when more powerful frequency domain measurements are required including modulation analysis, consider the keysight 89601a vector signal analyzer software ? debug your toughest designs with confidence (continued) i 2 c/spi serial trigger and decode (n5391a-1nl or option 007 on new scope purchases) given even further insights with protocol decode capability ? quickly move between physical and protocol layer information using the time-correlated tracking marker ? display protocol content using waveform symbols and the industrys first multi-tab protocol viewer ? the packets tab shows a high level view of the packet over time ? engineered for unmatched real-time measurement accuracy (continued) hardware accelerated differential and common mode math select the channel menu and enable differential mode to enable hardware accelerated math capability ? enjoy full channel functionality including infiniiscan triggering and jitter analysis ? save time, by using the hardware acceleration for even faster update rates with your differential and common mode math needs ?
08 | keysight | infiniium 90000 series oscilloscopes - data sheet debug your toughest designs with confidence (continued) ezjit analysis software (e2681a-1nl or option 002 on new scope purchases) quickly characterize and evaluate most commonly needed jitter measurements, including cycle-cycle, n-cycle, period, time-interval, error, setup and hold time, histograms, measurement trending and jitter spectrum ? includes advanced clock recovery options such as constant frequency and pll ? make measurements on repetitive or arbitrary data ? this application is supported on all models and is standard on dsa models ? for more information: www.keysight.com/find/ezjit conduct jitter analysis ? analyze jitter plus rj/dj separation ? ezjit complete ? engineered for unmatched real-time measurement accuracy (continued) ezjit complete analysis software (n8813a-1nl or option 070 on new scope purchases) ezjit complete includes all of the advanced jitter analysis capabilities of ezjit and ezjit plus, and adds advanced analysis of the vertical noise affecting the ones and zeros of your real-time eye ? decomposition of vertical noise provides key insight into degradation of your eye height ? in providing advanced decomposition of both horizontal jitter and vertical noise components of your signals, ezjit complete represents the most comprehensive analysis software available ? this application is supported on all models and is a standard on dsa models ? for more information: www.keysight.com/find/ezjitcomplete ezjit plus analysis software (n5400a-1nl or option 004 on new scope purchases ? to upgrade from ezjit to ezjit plus, order n5401a ? ) ezjit plus adds additional compliance views and an expanded measurement setup wizard to simplify and automate rj/dj separation for testing against industry standards ? this application is supported on all models and is standard on dsa models ? for more information: www.keysight.com/find/ezjitplus
09 | keysight | infiniium 90000 series oscilloscopes - data sheet identify signal integrity issues with infiniiscan zone C qualify triggering ? quickly characterize and correct for any input to your oscilloscope ? recover embedded clocks with serial data analysis (sda) ? engineered for unmatched real-time measurement accuracy (continued) debug your toughest designs with confidence (continued) infiniiscan event identification (n5414b-1nl or option 009 on new scope purchases) rapidly trigger on complex events and identify signal integrity issues ? this innovative software quickly scans through thousands of acquired waveform cycles and isolates anomalous signal behavior ? this application is supported on all models ? for more information: www.keysight.com/find/infiniiscan precisionprobe software (n2809a-1nl or option 001 on new scope purchases) make more accurate measurements independent of what probes or cables used ? keysights n2809a precisionprobe software characterizes and corrects for the loss in your specific cable or probe ? precisionprobe removes the uncertainty about the input connected to your oscilloscope by allowing you to see its characteristics in less than five minute ? precisionprobe gives you design and debug confidence by allowing you to quickly de-embed probe and cable loss to make more accurate measurements ? for more information: www.keysight.com/find/precisionprobe high-speed serial data analysis software (e2688a-1nl or option 003 on new scope purchases) quickly validate signal integrity for high-speed serial interfaces with embedded clocks ? recover embedded clocks synchronized with the analog waveform view ? build and validate eye diagrams ? the sda package also includes software-based bit-level triggering and decode for 8b/10b ? t his application is supported on all models and comes standard on dsa models ? for more information: www.keysight.com/find/sda
10 | keysight | infiniium 90000 series oscilloscopes - data sheet reduce receiver errors by opening tightly shut eyes ? control your applications remotely ? model channel effects including reflection ? engineered for unmatched real-time measurement accuracy (continued) debug your toughest designs with confidence (continued) infiniium remote programming interface (now standard on all infiniium scopes) operate your infiniium compliance and validation applications remotely using ? n et languages ? this application is supported on all models ? for more information: www.keysight.com/find/rpi serial data equalization (n5461a-1nl or option 012 on new scope purchases) measure at the pin and use equalization to see a virtual eye on the other side of an equalizer ? model equalization techniques such as dfe, ffe, and ctle ? this application is supported on all models ? for more information: www.keysight.com/find/sde infiniisim waveform transformation and de-embedding toolset (basic: n5465a-3nl or option 13 on new scope purchases ? advanced: n5465a-1nl or option 14 on new scope purchases ? ) the n5465a infiniisim waveform transformation toolset for infiniium series oscilloscopes provides the most flexible and accurate means to render waveforms anywhere in the digital serial link ? the configurable system modeling allows you to remove the deleterious effects of unwanted channel elements, simulate waveforms with channels models inserted, view waveforms in physically un-probable locations, compensate for loading of probes and other circuit elements, and do so simply and quickly on the real-time oscilloscope ? this application is supported on all models ? for more information: www.keysight.com/find/infiniisim
11 | keysight | infiniium 90000 series oscilloscopes - data sheet debug your toughest designs with confidence (continued) matlab data analysis software (option 061 or o62 on new scope purchases) matlab is a data analysis software environment and scripting language used by over 1,000,000 users in aerospace/defense, automotive, communications, electronics, and other applications ? matlab is now available directly from keysight as in instrument option with the purchase of your keysight 90000 series oscilloscope ? install matlab on your oscilloscope or remote pc to make customized measurements, design and apply your own filters to oscilloscope signals, graphically visualize signals in 2-d or 3-d plots, automate measurements, or build test applications ? purchase matlab with your keysight 90000 series oscilloscope to ensure version compatibility and so that your matlab software license is always available when you need it ? for more information: www.keysight.com/find/matlab_oscilloscopes signal equalization using user-defined function ? control your applications remotely ? quickly automate oscilloscope measurements ? engineered for unmatched real-time measurement accuracy (continued) user-defined function (n5430a-1nl or option 010 on new scope purchases) if we havent provided exactly what you need, use the n5430a user defined function software to create it yourself ? develop your own math functions or filters using matlab ? your custom functionality is seamlessly integrated into the infiniium 90000 menus and results are displayed on the scope screen ? this requires matlab (available as option 062) to be installed directly on the oscilloscope ? keysight is the only t&m manufacturer today that sells and supports matlab as its own product ? this application is supported on all models and requires matlab software (not included with udf) ? for more information: www.keysight.com/find/udf user-definable application (n5467b-1nl or option 040 on new scope purchases) rapidly develop your own automated measurements and tests ? this application provides the framework you need to quickly program and automate any single or set of measurements the oscilloscope can make ? the application also provides full control of other keysight instruments and html reporting capabilities ? for more information: www.keysight.com/find/uda
12 | keysight | infiniium 90000 series oscilloscopes - data sheet use vector signal analysis software to see fft-based spectrum analysis ? debug your toughest designs with confidence (continued) vector signal analysis software (89601a) expand the measurement capability of your scope with the 89601a vector signal analysis software ? this advanced dsp-based software takes the digitized signal data from the scope and displays fft-based spectrum analysis and wide-bandwidth digital modulation analysis for wireless communication signals such as w-cdma and cdma2000 ? and wireless networking signals such as 802 ? 1 1 wifi and 802 ? 1 6 wimax ? ? take advantage of the super-wide bandwidth of your scope to capture and evaluate radar signals ? for more information: www.keysight.com/find/vsa engineered for unmatched real-time measurement accuracy (continued) i 2 c/spi serial trigger and decode (n5391a-1nl or option 007 on new scope purchases) this application displays real-time time-aligned decode of i 2 c and spi packets ? view decode in waveform area or in protocol lister ? this application works on all models ? for more information: www.keysight.com/find/90000_i 2 c-spi trigger and view on-screen serial decode of i 2 c packets ? trigger on and decode jtag packets ? jtag (ieee 1149 ? 1 ) triggering and decode (n8817a-1nl or option 042 on new scope purchases) this application displays real-time time-aligned decode of jtag (ieee 1149 ? 1 ) tdi and tdo signals ? the application eliminates the difficult task of manually determining jtag tap controller states, instruction and data register decode, and flags error conditions ? the application includes scan chain description features including the ability to import ? b sdl files for each device and displays device names and opcodes in the protocol listing ? this application works on all models and can use any combination of scope or logic acquisition channels ? for more information: www.keysight.com/find/90000_jtag
13 | keysight | infiniium 90000 series oscilloscopes - data sheet trigger on and decode rs-232/uart transmission ? isolate signal integrity problems from logic-level coding errors on bidirectional serial data streams ? engineered for unmatched real-time measurement accuracy (continued) debug your toughest designs with confidence (continued) rs-232/uart serial decode and trigger (n5462a-1nl or option 015 on new scope purchases) this application eliminates the need to manually decode bus traffic ? using data captured on the scope channels, the application lets you easily view the information sent over an rs-232 ? display real-time time-aligned decode of transmit and receive lines ? this application works on all models ? for more information: www.keysight.com/find/90000_rs-232 n8805a-1nl usb 3 ? 0 p rotocol triggering and decode trigger on and view usb 3 ? 0 w ith the industrys first oscilloscope-based protocol analyzer with time-correlated views of physical layer and transaction layer errors ? the multi-tab protocol viewer includes correlation between the waveforms and the selected packet, enabling you to quickly move between the physical and protocol layer using the time-correlated tracking marker ? for more information: www.keysight.com/find/usb3decode
14 | keysight | infiniium 90000 series oscilloscopes - data sheet pass todays demanding compliance tests more quickly usb serial trigger and protocol viewer (n5464a-1nl or option 016 on new scope purchases) trigger on and quickly view usb 2 ? 0 p ackets, payload, header and detail information ? powerful time-correlated views of waveform and symbol, to the bit level, make it easy to isolate communication faults ? this application is supported on all models ? for more information: www.keysight.com/find/90000_usb_protocol_viewer trigger on and decode sas/sata serial packets ? trigger on and decode pcie serial packets ? trigger on and decode usb packets ? engineered for unmatched real-time measurement accuracy (continued) pci express serial trigger and protocol viewer (n5463a-1nl or option 017 on new scope purchases) this application provides protocol-level triggering and viewing of a pcie ? lane ? quickly view packets, payload, header, and detail information ? powerful time-correlated views of waveform, symbol, character, link and transaction layer packet data down to the bit level make it easy to isolate communication faults to logic or analog sources ? this application is supported on all 4 ghz and greater models ? for more information: www.keysight.com/find/90000_pci_protocol_viewer sata triggering and decode (n8801a-1nl or option 018 on new scope purchases) trigger on and view both protocol layer information and physical layer signal characteristics for sata 1 ? 5 g b/s, 3 ? 0 g b/s, and 6 ? 0 g b/s ? numerical decode values are automatically displayed and synchronizes below the capture signal or seen in protocol viewer ? this application works on all models ? for more information: www.keysight.com/find/n8801a
15 | keysight | infiniium 90000 series oscilloscopes - data sheet pass todays demanding compliance tests more quickly (continued) mipi ? d-phy sm trigger and decode (n8802a-1nl or option 019 on new scope purchases) this application eliminates the need to manually decode bus traffic ? using data captured on the scope, the application lets you easily view the information sent over mipi serial buses ? the application also enables software based protocol triggering ? this application is supported on all models 4 ghz bandwidth. for more information: www.keysight.com/find/n8802 a r trigger on and decode mipi packets ? validate ethernet compliance ? quickly verify and debug your pci express designs ? verify and debug your hdmi designs ? engineered for unmatched real-time measurement accuracy (continued) hdmi electrical performance validation and compliance software quickly verify and debug your high definition multi-media interface (hdmi) designs ? the n1080a fixture provides access to the compliance points for the electrical measurements required for the transmitter compliance testing ? switch matrix support is available by ordering n5399c-7nl (option 702 on new scope purchases). this application is support on all models 8 ghz bandwidth. for more information: www.keysight.com/find/n5399c pci express electrical performance validation and compliance software (n5393f or option 022 on new scope purchases) provides fast and easy way to verify and debug your pci express designs ? allows you to automatically execute pci express electrical checklist tests, and displays the results in a flexible report format ? ensures that your gen2 measurements will have absolute consistency with measurements made using the pci-sigs standalone sigtest software ? for more information: www.keysight.com/find/n5393d ethernet compliance testing (n5392b-3nl or option 021 on new scope purchases) perform a wide range of electrical tests for 10-, 100-, and 1000-base-t systems ? an n5395b or n5395c test fixture and n5396a jitter test cable speed compliance testing ? this application is supported on all models ? for more information: www.keysight.com/find/n5392b
16 | keysight | infiniium 90000 series oscilloscopes - data sheet pass todays demanding compliance tests more quickly (continued) energy efficient ethernet application (n5392b-1nl or option 060 on new scope purchases) industry leading compliance application that measures 10/100/1000bte energy efficient ethernet ieee 802 ? 3 az-2010 specification ? exclusive eee fixture allows you to measure all test modes at all speeds ? detailed reporting with clear measurements and pass/fail results ? this application is supported on all models ? for more information: www.keysight.com/find/eee sata 6g compliance test software (n5411b-1nl or option 038 on new scope purchases) rapidly validate and debug your sata 1 ? 5 gb/s (gen 1), 3 ? 0 g b/s (gen2) and 6 ? 0 g b/s (gen3) silicon, host bus adapter, port multiplier, high-density disk drive, solid-state disk drive or optical disk drive ? provides automated compliance test support for the i (internal), m(esata) and x(sas attachment) interfaces points, and displays the results in a flexible report format ? this application is supported on all models 12 ghz bandwidth. for more information: www.keysight.com/find/n5411b displayport 1 ? 2 c ompliance test software (u7232c-1nl or option 045 on new scope purchases) sets the benchmark for ease-of-use, and offers complete testing without compromise ? the software guides the user sequentially through the tasks ensuring minimal setup error, executes the tests specified by the standard and conveys the test information through a convenient software generated report ? the three modes of physical layer test allow for automated measurements based on the customizable configuration of compliance and characterization testing ? to make the test signal connection, the keysight w2641b displayport test point access adaptor completes the displayport source solution ? displayport 1 ? 2 a lso comes with switch matrix support to make multi-lane testing significantly easier by ordering u7232c-7nl (option 701 on new scope purchases) ? the application is supported on all models 8 ghz. for more information: www.keysight.com/find/u7232c characterize and evaluate the signal integrity of your energy efficient ethernet devices ? simplify the validation of sata designs ? full suite of displayport source tests ? engineered for unmatched real-time measurement accuracy (continued)
17 | keysight | infiniium 90000 series oscilloscopes - data sheet pass todays demanding compliance tests more quickly (continued) usb 2 ? 0 c ompliance test software (n5416a-1nl or option 029 on new scope purchases) quickly determine usb compliance with this usb-if recognized solution ? a setup wizard guides you through test selection and configuration ? this application is supported on all models ? for more information: www.keysight.com/find/n5416a check for usb compliance ? perform automated testing and margin analysis for xaui and xaui-derived specifications ? automatically execute d-phy electrical checklist tests for csi and dsi architectures ? engineered for unmatched real-time measurement accuracy (continued) mipi d-phy compliance test software (u7238c-1nl or option 035 on new scope purchases) automatically execute d-phy electrical checklist tests for csi and dsi architectures ? displays the results in a flexible report format ? mipi d-phy also supports the ability to test compliance with switches, making your testing faster, by ordering u7238c-7nl (option 703 on new scope purchases) ? the application is supported on all models ? for more information: www.keysight.com/find/d-phy_compliance xaui electrical validation with 10gbase-cx4, cpri, obsai, and serial rapidio support (n5431a-1nl or option 030 on new scope purchases) improve your efficiency by confirming that your devices conform to the xaui specifications as defined by the ieee 802 ? 3 -2005 10-gigabit ethernet specification ? provides support for the xaui-derived 10gbase-cx4 specification ? the application is supported on all models ? for more information: www.keysight.com/find/n5431a
18 | keysight | infiniium 90000 series oscilloscopes - data sheet pass todays demanding compliance tests more quickly (continued) 10gbase-t ethernet electrical conformance application for infiniium oscilloscopes (u7236a-1nl or option 036 on new scope purchases) takes care of the tedious task of instrument control and configures the oscilloscope, spectrum analyzer, or vector network analyzer as needed by each 10gbase-t test to provide rapid, accurate, and repeatable test execution ? the application is supported on all models ? for more information: www.keysight.com/find/10gbase-t automatically execute 10gbase-t ethernet physical-layer (phy) electrical tests ? validate and debug your usb 3 ? 0 s ilicon, host, hub or device ? test ddr memory ? comprehensive analysis that automates the complex measurements even when you are not there ? engineered for unmatched real-time measurement accuracy (continued) usb 3 ? 0 /3 ? 1 c ompliance test software (u7243b-3nl or option 041 on new scope purchases provides industry leading automated test support for usb 3 ? 0 p roducts and displays the test results in a comprehensive test report ? for best measurement accuracy use the keysight u7243b usb 3 ? 0 t ransmitter and receiver test fixtures ? keysights usb 3 ? 0 t est solution is designed from the ground up with the needs of the test engineer in mind ? the application is supported on models 12 ghz. for more information: www.keysight.com/find/usb3 ddr1 and lpddr/ddr2, lpddr2/ddr3, ddr4 and gddr5 compliance testing (options 031/032/033/058 on new scope purchases) quickly and easily evaluate and characterize your memory designs ? automated testingbased on jedec specifications saves time ? the application also includes additional debug and compliance capabilities ? this application is supported on all models ? however, the ddr technology you are using may dictate the minimal bandwidth required for your scope ? for more information: www.keysight.com/find/ddr sd uhs-i and sd uhs-ii card compliance testing (u7246a-1nl/n6461a-1nl) the sd card compliance test software gives you a fast, easy way to test, debug and characterize your sd designs up to 1 ? 5 g b/s in accordance with the sd specification ? the u7246a with 9000a and 90000a series oscilloscopes is a certified test tool for sd card phy electric tests with the sd association ? reference is available in sda official document: sd test tool information ver1 ? 0 s ep ? 13, 2010 ? for more information: www.keysight.com/find/u7246a www.keysight.com/find/n6461a
19 | keysight | infiniium 90000 series oscilloscopes - data sheet keysight infiniium portfolio keysights infiniium oscilloscope lineup includes bandwidths from 500 mhz to 63 ghz ? use the following selection guide to determine which best matches your specific needs ? infiniium real-time oscilloscopes feature the following: C worlds highest bandwidth on 4 channels in a single frame C industrys lowest noise floor C full precisionprobe compatibility s-series 90000a series 90000 x-series z-series available bandwidths up to 4 ghz 500 mhz, 1 ghz, 2.5 ghz, 4 ghz 2.5 ghz, 4 ghz 6 to 16 ghz 6 ghz (2 ch), 8 ghz (2 ch) 6 ghz, 8 ghz, 12 ghz, 13 ghz 13 ghz, 16 ghz 20 to 63 ghz 20 ghz, 25 ghz, 33 ghz 20 ghz, 25 ghz, 33 ghz, 50 ghz, 63 ghz max upgradable bandwidth 8 ghz 13 ghz 33 ghz 63 ghz sample rate (2-channel/4-channel) 10/20 gsa/s 40/40 gsa/s 80/40 gsa/s 160/80 gsa/s channel inputs and connector types 50 ? and 1 m?, bncs 50 ?, bncs 50 ?, 2.92 and 3.5 mm smas 50 ?, 1.85 mm, 2.4 mm, 2.92 and 3.5 mm, smas memory depth (standard/max) 50 m/1 gpts 20 m/2 gpts 20 m/2 gpts 50 m/2 gpts mso models yes no no no supported infiniimax probe families infiniimax 2 infiniimax 2 infiniimax 3 infiniimax 2 with adapter infiniimax 3 infiniimax 2 with adapter
20 | keysight | infiniium 90000 series oscilloscopes - data sheet infiniium 90000a series click on the icon at the bottom left of the infiniium screen to minimize the status and scales tab for full screen viewing ? maximize your viewing needs ? ever wanted to change the scale or offset of a function or waveform memory? if you have, you know that it requires multiple menus and key strokes ? in infiniium software version 2 ? 0 1 and later, you can now map functions and waveform memories to the front panel controls of the oscilloscope! starting with an 18-ghz, bnc-compatible connector, an ultra - l ow noise floor front end design using faraday cage technology ensures high signal integrity in its signal path ?
21 | keysight | infiniium 90000 series oscilloscopes - data sheet simply press the horizontal delay knob to set the delay value to zero ? a zoom button provides quick access to two screen zoom mode ? dedicated single acquisition button provide better control to capture an unique event ? customizable multipurpose key gives you any five automated measurements with a push of a button ? you can also configure this key to execute a script, print/save screen shots, save waveforms, or load a favorite setup ? measure section including a toggling marker button and a dedicated marker knob provides quick access to your marker control ? quick access to fine/vernier control by pressing the horizontal and vertical sensitivity knobs ? increase your productivity with a familiar infiniium graphical user interface, like your favorite drag - a nd - dr op measurement icons ? infiniiums analog-like front panel has a full set of controls color coded to the waveforms and measurements, making your tasks simple ? optional usb external dvd-rw drive allows you to install your favorite third - p arty software conveniently and can be used to back up your critical measurement data ? install third-party software packages on windows xp pro operating system such as excel, labview, keysight vee, matlab, anti-virus software, and more, to perform customized processing and automation of your oscilloscope or to make the scope compliant to the network environment of your company ? autoprobe interface completely configures your scope for use with the infiniimax probing system and previous-generation keysight active probes ?
22 | keysight | infiniium 90000 series oscilloscopes - data sheet infiniimax offers you the highest performance available for measuring differential and single - e nded signals, with flexible connectivity solutions for todays high- density ics and circuit boards ? infiniimax probes have fully characterized performance for all of their various probe heads ? this includes: C swept frequency response plot C common mode rejection versus frequency plot C impedance versus frequency plot C time-domain probe loading plot C time-domain probe tracking plot controlled impedance transmission lines in every probe head deliver full performance versus the performance limitations introduced by traditional wire accessories ? probe interface software allows you to save the calibration information for up to 10 different probe heads per channel and will automatically retrieve calibration data for a probe amplifier when attached to the scope ? high-input impedance active probes minimize loading, support differential measurements and dc offset, and can compensate for cable loss ? probe calibration software delivers the most accurate probe measurements and linear phase response and allows various probe combinations to be deskewed to the same reference time ? n5426a n5451a a flat frequency response over the entire probe bandwidth eliminates the distortion and frequency - d ependent loading effects that are present in probes that have an in-band resonance ? e2695a 8-ghz differential sma probe head allows you to connect two sma cables to make a differential measurement on a single scope channel ? e2679a 6 - g hz extremely small single-ended, solder - in probe heads for probing even the hardest - t o - r each single-ended signals ? n5425a 13 - g hz high - b andwidth solder - i n differential zif probe head and n5426a zif tip provides maximum bandwidth with the industrys first lead-free solder-in probe solution in an economical replaceable tip form factor ? n5451a 9 - ghz/5 - ghz long-wire zif tip provides a high - b andwidth economical replaceable solder-in tip with extra reach (9 ghz with 7 mm and 5 ghz with 11 mm wire) ? n5451a 9 - ghz/5 - ghz long-wire zif tip provides a high - b andwidth economical replaceable solder-in tip with extra reach (9 ghz with 7 mm and 5 ghz with 11 mm wire) ? e2695a 8-ghz differential sma probe head allows you to connect two sma cables to make a differential measurement on a single scope channel ? infiniimax ii: the worlds best high-speed probing system just keeps getting better
23 | keysight | infiniium 90000 series oscilloscopes - data sheet n5382a 13 - g hz high - ba ndwidth differential browser provides maximum bandwidth for hand- held or probe holder use ? variable spacing from 0 ? 2 t o 3 ? 3 mm ( 8 to 130 mills) ? e2675a 6 - g hz differential browser is the best choice for general-purpose trouble shooting of differential or single- ended signals with z-axis compliance and variable spacing from 0 ? 2 5 to 5 ? 8 0 mm (10 to 230 mills) ? e2676a 6 - g hz single-ended browser is the best choice for general-purpose probing of single-ended signals when the small size of the probe head is the primary consideration ? e2678a 12 - g hz differential socket probe head can be used to measure either differential or single - e nded signals via a plug-on socket connection ? n2880a in-line attenuator kit allows you to increase the dynamic range and the offset range of the infiniimax probe without affecting the bandwidth ? six different infiniimax probe amplifiers from 1 ? 5 g hz to 13 ghz are available for matching your probing solution to your performance and budget requirements ? the 1168/69a infiniimax ii amplifiers offer the highest bandwidth and the lowest noise floors ? the 1134/32/31/30a offer a more cost efficient solution and wider dynamic range ? n5380b 13 - g hz high - ba ndwidth differential sma probe head provides maximum bandwidth for sma - f ixtured differential pairs ? n5450b infiniimax extreme temperature extension cable provides extra reach into environmental chambers ? n2884a differential fine-wire probing tip infiniimax differential fine-wire probing tip is a high fidelity, high bandwidth solution for probing an active ic ? n2887a infiniimax soft touch pro probe adapter adapts from the keysight pro series (36 ch) soft touch connectorless logic analyzer foot print to the keysight infiniimax i & ii series probe amplifier input connectors ? n2881a dc blocking capacitors can be used to in series with the n2880a infiniimax in-line attenuators to block out unwanted dc components of the input signal up to 30v ? n2888a infiniimax soft touch half-channel probe adapter adapts from the keysight half-channel (18 ch) soft touch connectorless logic analyzer foot print to the keysight infiniimax i & ii series probe amplifier input connectors ?
24 | keysight | infiniium 90000 series oscilloscopes - data sheet infiniium 90000a series oscilloscopes performance characteristics vertical input channels four analog bandwidth (C3 db) 1, 5 90254a 90404a 90604a 90804a 91204a 91304a 2.5 ghz 4 ghz 6 ghz 8 ghz 12 ghz 12 ghz dsp enhanced bandwidth 4 91304a: 13-ghz real-time, user-selectable dsp enhanced bandwidth rise time/fall time 6 90254a 90404a 90604a 90804a 91204a 91304a 10 to 90% 140 ps 105 ps 70 ps 54 ps 35 ps 32 ps 20 to 80% 105 ps 79 ps 53 ps 38 ps 25 ps 23 ps input impedance 7 50 ?, 3% sensitivity 2 1 mv/div to 1 v/div input coupling dc vertical resolution 3 8 bits, 12 bits with averaging channel to channel isolation (any two channels with equal v/div settings dc to 3 ghz: 90804a/91204a/91304a: 60 db ( 1000:1) 90254a/90404a/90604a: 50 db ( 316:1) 3 to 8 ghz: 40 db ( 100:1) 8 ghz to bw: 35 db ( 56:1) dc gain accuracy 1 , 2 2% of full scale at full resolution channel scale ( 2.5% for 5mv/div) maximum input voltage 1 5 v offset range vertical sensitivity available offset 0 mv/div to 40 mv/div 0.4 v > 40 mv/div to 75 mv/div 0.9 v > 75 mv/div to 130 mv/div 1.6 v > 130 mv/div to 240 mv/div 3.0 v > 240 mv/div 4.0 v offset accuracy 1 , 2 3.5 v: (2% of channel offset + 1% of full scale) + 1 mv > 3.5 v: (2% of channel offset + 1% of full scale) dynamic range 4 div from center screen dc voltage measurement accuracy 1 , 2 dual cursor: [(dc gain accuracy) + (resolution)] single cursor: [(dc gain accuracy) + (offset accuracy) + (resolution/2)] rms noise floor (scope only) volts/div 90254a 90404a 90604a 90804a 91204a 91304a 5 mv 153 v 199 v 259 v 322 v 435 v 467 v 10 mv 183 v 232 v 295 v 358 v 483 v 536 v 20 mv 275 v 342 v 424 v 498 v 650 v 758 v 50 mv 645 v 799 v 985 v 1.15 mv 1.45 mv 1.73 mv 100 mv 1.27 mv 1.56 mv 1.92 mv 2.22 mv 2.80 mv 3.37 mv 200 mv 2.47 mv 3.03 mv 3.71 mv 4.28 mv 5.41 mv 6.58 mv 500 mv 6.48 mv 8.00 mv 9.91 mv 11.5 mv 14.7 mv 17. 4 mv 1 v 12.5 mv 15.6 mv 19.2 mv 22.3 mv 28.5 mv 34.1 mv hqrwhvzduudqwhgvshflofdwlrqvdoorwkhuvduhwslfdo6shflofdwlrqvduhydolgdiwhudplqxwhzdupxsshulrgdqgiurpdqqxdofdoleudwlrq temperature ? ) xoovfdohlvghoqhgdvyhuwlfdoglylvlrqv0djqlofdwlrqlvxvhgehorzp9gly%horzp9glyixoovfdohlvghoqhgdvp97khpdmruvfdohvhwwlqjvduh 5 mv, 10 mv, 20 mv, 50 mv, 100 mv, 200 mv, 500 mv, 1 v ? 3 ? v ertical resolution for 8 bits = 0 ? 4 % of full scale, for 12 bits = 0 ? 0 24% of full scale ? 4 ? 1 3 ghz dsp enhanced bandwidth not applicable at 5 mv/div ? ) )7dpsolwxghuhdglqjvduhdiihfwhgevfrshdqgsurehedqgzlgwkolplwdwlrqvdqglqsxwdpsolohuvurooriihjg%urooriidwvshflohgedqgzlgwkri scope/probe) ? 6 ? t he fft signal to noise ratio varies with volts/division setting, memory depth and use of time or frequency averaging ? d 1 rlvhlvwkhglvsodhgqrlvhprru6ohz5dwhlvwkhglvsodhgvohzudwhriwkhvljqdodwwkhwkuhvkrogfurvvlqjv6dpsohudwh pdvlqlqwhusrodwlrq enabled ? e 0 hdvxuhphqwwkuhvkrog ohgyrowdjhdwohyho f 7 lphudqjhvv 8 ? v alues represent time error between two edges on a single channel ? standard deviation value refers to the standard deviation of 256 consecutive phdvxuhphqwvshuiruphgxvlqjdqlqglylgxdolqvwuxphqw5hdglqjlvwkhglvsodhg70phdvxuhphqwydoxh7lph6fdohfflvwkhrvfloorvfrshjvvshflohg time scale accuracy ? , qwhuqdohgjhwuljjhuprgh7uljjhuwkuhvkrog ohgyrowdjhdwohyho7khvohzudwhlqghshqghqwydoxhlqwkhirupxoduhsuhvhqwvwkhwudglwlrqdowuljjhu jit ter ? 5htxluhv2swlrqxvhughoqhgixqfwlrq 11 ? 11 ? 8 g hz analog bandwidth at 5 mv/div for dso91304a and dso91204a models ? 12 ? typically triggers as low as 5 mv/div sensitivity ?
25 | keysight | infiniium 90000 series oscilloscopes - data sheet infiniium 90000a series oscilloscopes (continued) performance characteristics (continued) vertical (continued) rms noise floor (scope with probe) volts/div 90254a with 1131a 90404a with 1132a 90604a with 1134a 90804a with 1168a 91204a with 1169a 91304a with 1169a 20 mv 3.2 mv 3.5 mv 4.0 mv 2.2 mv 2.5 mv 2.7 mv 50 mv 3.3 mv 3.6 mv 4.0 mv 2.3 mv 2.8 mv 3.1 mv 100 mv 3.4 mv 3.8 mv 4.3 mv 2.9 mv 3.5 mv 4.2 mv 200 mv 4.0 mv 4.6 mv 5.3 mv 4.7 mv 5.9 mv 7.5 mv 500 mv 7.1 mv 8.6 mv 10 mv 12 mv 15 mv 19 mv 1 v 13 mv 16 mv 19 mv 23 mv 28 mv 37 mv horizontal main timebase range 5 ps/div to 200 s/div real-time main timebase delay range C200 s to 200 s real-time zoom timebase range 1 ps/div to current main time scale setting time scale accuracy 1, 8 (0.4 ppm initial + 0.5 ppm/year aging) oscilloscope channel de-skew range 25 s range, 100 fs resolution intrinsic jitter 6 (sample clock jitter) acquired time range or delta-time delta-time interval internal reference external reference < 100 s (10 s/div) 0.5 ps rms 0.5 ps rms 1 ms (100 s/div) 1.0 ps rms 1.0 ps rms 10 ms (1 ms/div) 1.2 ps rms 1.2 ps rms 100 ms (10 ms/div) 5.0 ps rms 1.2 ps rms inter-channel intrinsic jitter 3 550 fs rms inter-channel skew drift 3, 7 < 500 fs rms jitter measurement floor 2 time interval error () 2 2  intr insi c jitter   noise floor slew rate ?? + ?? ?? period jitter () 2 2  2  noise floor intrinsicj itte r slew rate ?? + ?? ?? i cycle-cycle/n-cycle jitter () 2 2  3  noise floor intrinsi cj itte r slew rate ?? + ?? ?? i inter-channel jitter measurement floor 2, 3, 5 () () 22 2    12  time interval time interval inte rc hannel erro re dge erro re dge intrinsic jitter ?? ?? ?? ++ ?? ?? ?? ?? ?? ?? 1 ? hqrwhvzduudqwhgvshflilfdwlrqvdoorwkhuvduhwslfdo6shflilfdwlrqvduhydolgdiwhudplqxwhzdupxsshulrgdqg&iurpilupzduhfdoleudwlrq temperature ? 2 ? sample rate at maximum ? noise and slew rate determined at fixed-voltage measurement threshold, near middle of signal ? displayed signal not vertically folsshg6ohzudwhrivlqhzdyh shdnvljqdodpsolwxghvvvivohzudwhriidvwvwhsa wrulvhwlph 3 ? intra-channel = both edges on the same channel, inter-channel = two edges on different channels ? time interval error(edge1) = time-interval error measurement floor of first edge, time interval error(edge2) = time-interval error measurement floor of second edge ? 4 ? reading is the displayed delta time measurement accuracy measurement value ? do not double the listed time scale accuracy value in delta time measurement accuracy formula ? 5 ? scope channels and signal interconnect de-skewed prior to measurement ? 6 ? external timebase reference values measured using a wenzel 501-04608a 10 mhz reference ? intrinsic jitter value depends on acquisition time range for time interval error formula and depends on delta-time between edges for all two-edge formulas ? 7 ? 6nhzehwzhhqfkdqqhovfdxvhge&whpshudwxuhfkdqjh 8 ? initial = immediately after factory or user calibration ?
26 | keysight | infiniium 90000 series oscilloscopes - data sheet infiniium 90000a series oscilloscopes (continued) performance characteristics (continued) horizontal (continued) delta time measurement accuracy 2, 3, 4, 5 intra-channel no averaging () () 22     12  time interval time interval time scale erro re dge erro re dge accuracy ?? ?? ?? ?? ?? ?? delta time ?? ?? ?? ?? ?? ?? ?? ?? 5  + + ?? ?? ?? ?? ?? ii 256 averages () () 22     5   12  16 time interval time interval time scale erro re dge erro re dge accuracy ?? ?? ?? ?? ?? ?? ?? ii delta time ?? ?? + + ?? ?? ?? ?? ?? ?? ?? ?? ?? ?? inter-channel no averaging () () 22 2    5 12       time interval time interval inte rc hannel erro re dge erro re dge intrinsicj itte r time scale accuracy ?? ?? ?? ?? ++ ?? ?? ?? ?? ?? ?? ?? ?? ? ++ ?? ?? ?? ?? ? i i ? ? ?? ?? ?? ?? ?? ?? ?? ?? ?? ?? inte rc hannel delta time skew drif t 256 averages () () 22   5  12 16     time interval time interval erro re dge erro re dge time scale accuracy ?? ?? ++ ?? ?? ?? ?? i i ?? ?? ?? ?? ??   ?? ?? ?? ?? ++ ?? ?? ?? ?? ?? ?? ?? ?? ?? ?? inte rc hannel delta time skew drift 1 ? denotes warranted specifications, all others are typical. specifications are valid after a 30-minute warm-up period and 5 c from firmware calibration temperature ? 2 ? sample rate at maximum ? noise and slew rate determined at fixed-voltage measurement threshold, near middle of signal ? displayed signal not vertically clipped. slew rate of sine wave = (peak signal amplitude) ? 2 ? ? f, slew rate of fast step ~= (10 to 90% rise time). 3 ? intra-channel = both edges on the same channel, inter-channel = two edges on different channels ? time interval error(edge1) = time-interval error measurement floor of first edge, time interval error(edge2) = time-interval error measurement floor of second edge ? 4 ? reading is the displayed delta time measurement accuracy measurement value ? do not double the listed time scale accuracy value in delta time measurement accuracy formula ? 5 ? scope channels and signal interconnect de-skewed prior to measurement ? 6 ? external timebase reference values measured using a wenzel 501-04608a 10 mhz reference ? intrinsic jitter value depends on acquisition time range for time interval error formula and depends on delta-time between edges for all two-edge formulas ? 7 ? skew between channels caused by 5 oc temperature change. 8 ? initial = immediately after factory or user calibration ?
27 | keysight | infiniium 90000 series oscilloscopes - data sheet infiniium 90000a series oscilloscopes (continued) performance characteristics (continued) acquisition maximum real-time sample rate 91304a/91204a/90804a: 40 gsa/s (4 channels simultaneously) 90604a/90404a/90254a: 20 gsa/s (4 channels simultaneously) maximum acquired time at highest real-time resolution 91304a/91204a/90804a 90604a/90404a/90254a resolution 25 ps (40 gsa/s) 50 ps (20 gsa/s) standard 0.5 ms 1.0 ms option 50m 1.25 ms 2.5 ms option 100 2.5 ms 5.0 ms option 200 5.0 ms 10.0 ms option 500 12.5 ms 25.0 ms option 01g 25.0 ms 50.0 ms option 02g 50.0 ms 100.0 ms data transfer speed gigabit ethernet samples: 1 k 64 k 1 m 16 m 32 m 128 m msa/s (word): 0.1 1.88 9.25 12.00 12.80 12.80 msa/s (byte): 0.11 1.88 12.60 19.70 20.30 22.00 usb 2.0 hi-speed (device) samples: 1 k 64 k 1 m 16 m 32 m 128 m msa/s (word): 0.11 1.88 8.34 8.55 9.07 11.38 msa/s (byte): 0.11 1.88 11.60 14.40 14.90 18.10 sampling modes real-time successive single-shot acquisitions real-time with averaging selectable from 2 to 65534 real-time with peak detect 91304a/91204a/90804a: 40 gsa/s 90604a/90404a/90254a: 20 gsa/s real-time with hi resolution real-time boxcar averaging reduces random noise and increases resolution equivalent-time resolution: 338 fs full bandwidth on all 4 channels, 262,144 sample points maximum memory segmented memory captures bursting signals at maximum sample rate without consuming memory during periods of inactivity minimum intersegment time: 91304a / 91204a / 90804a: 2.7 s 90604a / 90404a / 90254a: 2.5 s (the time between the end of the previous acquisition and the beginning of the next acquisition) roll mode scrolls sequential waveform points across the display in a right-to-left rolling motion. works at sample rates up to 10 msa/s with a maximum record length of 40ms. filters sin(x)/x interpolation on/off selectable fir digital filter. digital signal processing adds points between acquired data points to enhance measurement accuracy and waveform display quality. memory depth: 20 m 50 m 100 m 200 m 500 m 1 g 2 g max # of segments: 4096 8192 16384 32768 65536 131072 131072
28 | keysight | infiniium 90000 series oscilloscopes - data sheet performance characteristics (continued) hardware trigger sensitivity 2 91304a/91204a/90804a: internal low 2 : 2.0 div p-p 0 to 5 ghz internal high 2 : 0.3 div p-p 0 to 4 ghz, 1.0 div p-p 4 to 7.5 ghz 90604a/90404a/90254a 12 : internal low 2 : 2.0 div p-p 0 to 5 ghz internal high 2 : 0.3 div p-p 0 to 3 ghz, 1.0 div p-p 3 to 5 ghz auxiliary: dc to 100 mhz: 200 mv p-p into 50 ? 100 mhz to 1 ghz: 500 mv p-p into 50 ? level range internal 4 div from center screen or 4 volts, whichever is smallest auxiliary 5 v, also limit input signal to 5 v sweep modes auto, triggered, single display jitter (trigger jitter) 2, 3, 4 620 fs rms trigger sources channel 1, channel 2, channel 3, channel 4, aux, and line trigger modes edge triggers on a specified slope (rising, falling or alternating between rising and falling) and voltage level on any channel or auxiliary trigger. edge transition trigger on rising or falling edges that cross two voltage levels in > or < the amount of time specified. edge transition setting from 250 ps. edge then edge (time) the trigger is qualified by an edge. after a specified time delay between 10 ns to 10 s, a rising or falling edge on any one selected input will generate the trigger. edge then edge (event) the trigger is qualified by an edge. after a specified delay between 1 to 16,000,000 rising or falling edges, another rising or falling edge on any one selected input will generate the trigger. glitch triggers on glitches narrower than the other pulses in your waveform by specifying a width less than your narrowest pulse and a polarity. triggers on glitches as narrow as 125 ps. glitch range settings: < 250 ps to < 10 s. line triggers on the line voltage powering the oscilloscope. pulse width trigger on a pulse that is wider or narrower than the other pulses in your waveform by specifying a pulse width and a polarity. triggers on pulse widths as narrow as 125 ps. pulse width range settings: 250 ps to 10 s. trigger point can be end of pulse or time out. runt triggers on a pulse that crosses one threshold but fails to cross a second threshold before crossing the first again. can be time qualified with minimum setting of 250 ps. timeout trigger when a channel stays high, low, or unchanged for too long. timeout setting: from 250 ps to 10 s. pattern/pulse range triggers when a specified logical combination of the channels is entered, exited, present for a specified period of time or is within a specified time range or times out. each channel can have a value of high (h), low (l) or dont care (x). state pattern trigger clocked by the rising, falling or alternating between rising and falling edge of one channel. setup/hold triggers on setup, hold, or setup and hold violations in your circuit. requires a clock and data signal on any two inputs (except aux or line) channels as trigger sources. setup and/or hold time must then be specified. 1 ? hqrwhvzduudqwhgvshflilfdwlrqvdoorwkhuvduhwslfdo6shflilfdwlrqvduhydolgdiwhudplqxwhzdupxsshulrgdqg&iurpilupzduhfdoleudwlrq temperature ? 2 ? internal edge trigger mode with jitterfree correction ? value depends on scope settings and trigger signal characteristics, and is equal to tie value expressed in the formula above using the minimum time scale accuracy value ? 3 ? value shown represents typical display jitter for 91204a at 100 mv/div triggering on 500 mvpp 6 ghz sin wave signal ? 4 ? sample rate at maximum ? noise and slew rate determined at fixed-voltage trigger threshold, near middle of signal ? displayed signal not vertically clipped ? infiniium 90000a series oscilloscopes (continued)
29 | keysight | infiniium 90000 series oscilloscopes - data sheet performance characteristics (continued) hardware trigger (continued) trigger modes (continued) window triggers on an event associated with a window defined by two-user adjustable thresholds. event can be window entered, exited, inside (time qualified), or outside (time qualified) voltage range. trigger point can be cross window boundary or time out. time qualify range: from 250 ps to 10 s. video triggers from negative sync composite video, field 1, field 2, or alternating fields for interlaced systems, any field, specific line, or any line for interlaced or non-interlaced systems. supports ntsc, pal-m (525/60), pal, secam (625/50), edtv (480p/60), edtv (576p/50), hdtv (720p/60), hdtv (720p/50), hdtv (1080i/60), hdtv (1080i/50), hdtv (1080p/60), hdtv (1080p/50), hdtv (1080p/30), hdtv (1080p/25), hdtv (1080p/24), and user-defined formats. trigger sequences three stage trigger sequences including two - s tage hardware (find event (a) and trigger event (b)) and one - s tage infiniiscan software trigger. supports all hardware trigger modes except edge then edge and video, and all infiniiscan software trigger modes. supports delay (by time) and reset (by time or event) between two hardware sequences. the minimum latency between find event (a) and trigger event (b) is 3 ns. trigger qualification and qualifier single or multiple channels may be logically qualified with any other trigger mode trigger holdoff range 100 ns to 10 s trigger actions specify an action to occur and the frequency of the action when a trigger condition occurs. actions include e-mail on trigger and execute multipurpose user setting. trigger shortcuts provides easy shortcuts to all trigger features software trigger (requires infiniiscan event identification software option 009) trigger modes generic serial software triggers on nrz-encoded data up to 8.0 gbps, up to 80 - b it pattern. support multiple clock data recovery methods including constant frequency, 1st - o rder pll, 2nd-order pll, explicit clock, explicit 1st-order pll, explicit 2nd-order pll, fibre channel, flexray receiver, flexray transmitter (requires e2688a except for the constant frequency clock data recovery mode). measurement limit software triggers on the results of the measurement values. for example, when the pulse width measurement is turned on, infiniiscan measurement software trigger triggers on a glitch as narrow as 75 ps. when the time interval error (tie) is measured, infiniiscan can trigger on a specific tie value. non-monotonic edge software triggers on the non-monotonic edge. the non-monotonic edge is specified by setting a hysteresis value. runt software triggers on a pulse that crosses one threshold but fails to cross a second threshold before crossing the first again. unlike hardware runt trigger, infiniiscan runt trigger can be further qualified via a hysteresis value. zone qualify software triggers on the user defined zones on screen. zones can be specified as either must intersect or must not intersect. up to four zones can be defined. infiniium 90000a series oscilloscopes (continued)
30 | keysight | infiniium 90000 series oscilloscopes - data sheet performance characteristics (continued) measurements and math maximum measurement update rate > 42,000 measurement/sec (one measurement turned on) > 122,000 measurement/sec/measurement (five measurements turned on) measurement modes standard, measure all edges mode waveform measurements voltage peak to peak, minimum, maximum, average, rms, amplitude, base, top, overshoot, preshoot, upper, middle, lower time rise time, fall time, period, frequency, positive width, negative width, duty cycle, burst width, tmin, tmax, tvolt, setup time (requires option 002, 004, or 070 standard on dsa models), hold time (requires option 002, 004, or 070 standard on dsa models), channel-to-channel delta time, channel - t o-channel phase mixed area, slew rate frequency domain fft frequency, fft magnitude, fft delta frequency, fft delta magnitude level qualifcation any channels that are not involved in a measurement can be used to level-qualify all timing measurements eye-diagram measurements eye height, eye width, eye jitter, crossing percentage, q factor, and duty-cycle distortion jitter analysis measurements fast eye rendering increases speed of the eye diagram rendering clock requires option 002 (or e2681a), 004 (or n5400a), or 070 (or n8823a). standard on dsa series. data time interval error (tie) data with tie band, high, low-pass filter, data rate, unit interval, clock recovery rate, burst time, burst period, burst interval timing two sources: setup time, hold time, phase, advanced one source: period, frequency, + width, width, duty cycle, burst width, rise time, fall time, slew rate statistics displays the current, mean, minimum, maximum, range (max-min), standard deviation, number of measurements value for the displayed automatic measurements histograms source waveform or measurement 13 orientation vertical (for timing and jitter measurements) or horizontal (noise and amplitude change) modes, regions are defined using waveform markers measurements mean, standard deviation, mean 1, 2, and 3 sigma, median, mode, peak-to-peak, min, max, total hits, peak (area of most hits), x scale hits, and x offset hits hqrwhvzduudqwhgvshflofdwlrqvdoorwkhuvduhwslfdo6shflofdwlrqvduhydolgdiwhudplqxwhzdupxsshulrgdqgiurpdqqxdofdoleudwlrq temperature ? ) xoovfdohlvghoqhgdvyhuwlfdoglylvlrqv0djqlofdwlrqlvxvhgehorzp9gly%horzp9glyixoovfdohlvghoqhgdvp97khpdmruvfdohvhwwlqjvduh 5 mv, 10 mv, 20 mv, 50 mv, 100 mv, 200 mv, 500 mv, 1 v ? 3 ? v ertical resolution for 8 bits = 0 ? 4 % of full scale, for 12 bits = 0 ? 0 24% of full scale ? 4 ? 1 3 ghz dsp enhanced bandwidth not applicable at 5 mv/div ? ) )7dpsolwxghuhdglqjvduhdiihfwhgevfrshdqgsurehedqgzlgwkolplwdwlrqvdqglqsxwdpsolohuvurooriihjg%urooriidwvshflohgedqgzlgwkrivfrshsureh 6 ? t he fft signal to noise ratio varies with volts/division setting, memory depth and use of time or frequency averaging ? d 1 rlvhlvwkhglvsodhgqrlvhprru6ohz5dwhlvwkhglvsodhgvohzudwhriwkhvljqdodwwkhwkuhvkrogfurvvlqjv6dpsohudwh pdvlqlqwhusrodwlrq enabled ? e 0 hdvxuhphqwwkuhvkrog ohgyrowdjhdwohyho f 7 lphudqjhvv 8 ? v alues represent time error between two edges on a single channel ? standard deviation value refers to the standard deviation of 256 consecutive phdvxuhphqwvshuiruphgxvlqjdqlqglylgxdolqvwuxphqw5hdglqjlvwkhglvsodhg70phdvxuhphqwydoxh7lph6fdohfflvwkhrvfloorvfrshjvvshflohg time scale accuracy ? , qwhuqdohgjhwuljjhuprgh7uljjhuwkuhvkrog ohgyrowdjhdwohyho7khvohzudwhlqghshqghqwydoxhlqwkhirupxoduhsuhvhqwvwkhwudglwlrqdowuljjhu jit ter ? 5 htxluhv2swlrqxvhughoqhgixqfwlrq 11 ? 1 1 ? 8 g hz analog bandwidth at 5 mv/div for dso91304a and dso91204a models ? 12 ? t ypically triggers as low as 5 mv/div sensitivity ? 13 ? m easurement histograms require ezjit license ? infiniium 90000a series oscilloscopes (continued)
31 | keysight | infiniium 90000 series oscilloscopes - data sheet performance characteristics (continued) measurements and math (continued) mask testing allows pass/fail testing to user-defined or keysight-supplied waveform templates. automask lets you create a mask template from a captured waveform and define a tolerance range in time/voltage or screen divisions. test modes (run until) include test forever, test to specified time or event limit, and stop on failure. executes multipurpose user setting on failure. unfold real time eye feature will allow individual bit errors to be observed by unfolding a real time eye when clock recovery is on. communications mask test kit option provides a set of itu-t g.703, ansi t1.102, and ieee 802.3 industry-standard masks for compliance testing. waveform math number of functions 16 hardware accelerated math differential and common mode operators absolute value, add, amplitude modulation, average, butterworth 10 , common mode, delay, differentiate, divide, fft magnitude, fft phase, fir 10 , high pass filter, histogram (measurement), horizontal gate, integrate, invert, lfe 10 , low pass filter (4th - o rder bessel thompson filter), magnify, max, measurement trend, min, multiply, rt eye 10 , smoothing, sqrtsumofsquare 10 , square, square root, subtract, versus, and optional user defined function (option 010) fft frequency range 5 dc up to 20 ghz (at 40 gsa/s) or 10 ghz (at 20 gsa/s) frequency resolution sample rate/memory depth = resolution best resolution at 91304a/91204a/90804a: 800 hz maximum sample rate 90604a/90404a/90254a: 400 hz frequency accuracy (1/2 frequency resolution) + (1 x 10- 6 )(signal frequency) hqrwhvzduudqwhgvshflofdwlrqvdoorwkhuvduhwslfdo6shflofdwlrqvduhydolgdiwhudplqxwhzdupxsshulrgdqgiurpdqqxdofdoleudwlrq temperature ? ) xoovfdohlvghoqhgdvyhuwlfdoglylvlrqv0djqlofdwlrqlvxvhgehorzp9gly%horzp9glyixoovfdohlvghoqhgdvp97khpdmruvfdohvhwwlqjvduh 5 mv, 10 mv, 20 mv, 50 mv, 100 mv, 200 mv, 500 mv, 1 v ? 3 ? v ertical resolution for 8 bits = 0 ? 4 % of full scale, for 12 bits = 0 ? 0 24% of full scale ? 4 ? 1 3 ghz dsp enhanced bandwidth not applicable at 5 mv/div ? ) )7dpsolwxghuhdglqjvduhdiihfwhgevfrshdqgsurehedqgzlgwkolplwdwlrqvdqglqsxwdpsolohuvurooriihjg%urooriidwvshflohgedqgzlgwkrivfrshsureh 6 ? t he fft signal to noise ratio varies with volts/division setting, memory depth and use of time or frequency averaging ? d 1 rlvhlvwkhglvsodhgqrlvhprru6ohz5dwhlvwkhglvsodhgvohzudwhriwkhvljqdodwwkhwkuhvkrogfurvvlqjv6dpsohudwh pdvlqlqwhusrodwlrq enabled ? e 0 hdvxuhphqwwkuhvkrog ohgyrowdjhdwohyho f 7 lphudqjhvv 8 ? v alues represent time error between two edges on a single channel ? standard deviation value refers to the standard deviation of 256 consecutive phdvxuhphqwvshuiruphgxvlqjdqlqglylgxdolqvwuxphqw5hdglqjlvwkhglvsodhg70phdvxuhphqwydoxh7lph6fdohfflvwkhrvfloorvfrshjvvshflohg time scale accuracy ? , qwhuqdohgjhwuljjhuprgh7uljjhuwkuhvkrog ohgyrowdjhdwohyho7khvohzudwhlqghshqghqwydoxhlqwkhirupxoduhsuhvhqwvwkhwudglwlrqdowuljjhu jit ter ? 5 htxluhv2swlrqxvhughoqhgixqfwlrq 11 ? 1 1 ? 8 g hz analog bandwidth at 5 mv/div for dso91304a and dso91204a models ? 12 ? t ypically triggers as low as 5 mv/div sensitivity ? 13 ? m easurement histograms require ezjit license ? infiniium 90000a series oscilloscopes (continued)
32 | keysight | infiniium 90000 series oscilloscopes - data sheet performance characteristics (continued) measurements and math (continued) fft (continued) signal-to-noise ratio 6 60 db to > 100 db depending on settings window modes hanning, flattop, rectangular, blackman-harris measurement modes automatic measurements measure menu access to all measurements, ten measurements can be displayed simultaneously multipurpose front-panel button activates ten pre-selected or ten user-defined automatic measurements drag-and-drop measurement toolbar measurement toolbar with common measurement icons that can be dragged and dropped onto the displayed waveforms snapshot takes 29 snap shot measurements (customizable) marker modes manual markers, track waveform data, track measurements display display 12.1 - in ch color xga tft-lcd with touch screen intensity grayscale 256 - l evel intensity - g raded display resolution xga 1024 pixels horizontally x 768 pixels vertically annotation up to 12 labels, with up to 100 characters each, can be inserted into the waveform area grids up to 16 grids each with 8 bit vertical resolution waveform styles connected dots, dots, infinite persistence, color graded infinite persistence. includes up to 256 levels of intensity - g raded waveforms. waveform update rate maximum waveform update > 400,000 waveforms per second (when in the segment memory mode) computer system and peripherals, i/o ports computer system and peripherals operating system windows 7 embedded standard cpu intel core 2 duo 3.06 ghz pc system memory 4 gb ddr2 (standard) drives 250 - g b internal hard drive optional removable hard drive (option 801) optional usb external dvd-rw drive (option 820) peripherals logitech optical usb mouse, compact usb keyboard and stylus supplied. all infiniium models support any windows-compatible input device with a serial, ps/2 or usb interface. file types waveforms (supported max memory size) compressed internal format (*.wfm (200 mpts)), comma - s eparated values (*.csv (1 gpts)), tab separated values (*.tsv (1 gpts)), public binary format (.bin (500 mpts)), y value files (*.txt (1 gpts)), hierarchal data file (*.hf5(1 gpts)), images bmp, png, tiff, gif or jpeg i/o ports lan rj-45 connector, supports 10base-t, 100base-t, and 1000base-t. enables web-enabled remote control, e-mail on trigger or demand, data/file transfers and network printing (vxi-11). recommended web remote control tool: ultra vnc ( http://www.ultravnc.com/ ). 1 ? hqrwhvzduudqwhgvshflilfdwlrqvdoorwkhuvduhwslfdo6shflilfdwlrqvduhydolgdiwhudplqxwhzdupxsshulrgdqg&iurpdqqxdofdoleudwlrq temperature ? 2 ? full scale is defined as 8 vertical divisions ? magnification is used below 5 mv/div ? below 5 mv/div, full-scale is defined as 40 mv ? the major scale settings are 5 mv, 10 mv, 20 mv, 50 mv, 100 mv, 200 mv, 500 mv, 1 v ? 3 ? vertical resolution for 8 bits = 0 ? 4 % of full scale, for 12 bits = 0 ? 0 24% of full scale ? 4 ? 13 ghz dsp enhanced bandwidth not applicable at 5 mv/div ? 5 ? fft amplitude readings are affected by scope and probe bandwidth limitations and input amplifiers roll-off (e ? g ? 3 db roll-off at specified bandwidth of scope/probe) ? 6 ? the fft signal to noise ratio varies with volts/division setting, memory depth and use of time or frequency averaging ? infiniium 90000a series oscilloscopes (continued)
33 | keysight | infiniium 90000 series oscilloscopes - data sheet performance characteristics (continued) computer system and peripherals, i/o ports (continued) i/o ports (continued) pci express pci express x4 link, enabled by sockets (optional- option 823) gpib ieee 488.2, fully programmable (optional C option 805) rs-232 (serial) com1, printer and pointing device support parallel centronics printer port ps/2 two ports. supports ps/2 pointing and input devices. usb 2.0 hi-speed (host) three usb 2.0 hi-speed host ports on front panel plus four usb 2.0 hi-speed host ports on rear panel usb 2.0 hi-speed (device) one usb 2.0 hi-speed device port on rear panel that enables usb instrument control dual-monitor video output 15 pin xga (1024x768), full color output of scope waveform display or dual monitor video output auxiliary output dc ( 2.4 v); square wave (~715 hz and ~456 mhz); trigger output (255 mv p-p into 50) trigger output 5 v 50 ? back-terminated time base reference output 10 mhz filtered sine wave with all harmonics C40 dbc. amplitude into 50 ?: 800 mv p-p to 1.26 v p-p (4 dbm 2 db) if derived from internal reference. tracks external reference input amplitude 1 db if applied and selected. time base reference input must be 10 mhz, input z0 = 50 ?. minimum 500 mv p-p (C2 dbm), maximum 2.0 v p-p (+10 dbm). lxi compliance functional class c general characteristics temperature operating: 5 c to +40 c; non-operating: C40 c to +65 c humidity operating: up to 95% relative humidity (non-condensing) at +40 c; non-operating: up to 90% relative humidity at +65 c altitude operating: up to 4,000 meters (12,000 feet); non-operating: up to 15,300 meters (50,000 feet) vibration operating random 0.21 g(rms), non-operating random 2.0 g(rms), swept sins (0.50 g(rms)) power 100 to 240 vac at 50/60 hz; maximum input power 800 w weight net: 20 kg (44 lbs.) shipping: 27.4 kg (60 lbs.) dimensions (excluding handle) height: 283 mm (11.13 inch); width: 432 mm (17.02 inch); depth: 506 mm (19.91 inch) safety meets iec 61010-1 +a2, csa certified to c22.2 no.1010.1, self-certified to ul 3111 1. d enotes warranted specifcations, all others are typical. specifcations are valid after a 30-minute warm-up period, and 5 c from annual calibration temperature ? ) xoovfdohlvghoqhgdvyhuwlfdoglylvlrqv0djqlofdwlrqlvxvhgehorzp9gly%horzp9glyixoovfdohlvghoqhgdvp97khpdmruvfdohvhwwlqjvduh 5 mv, 10 mv, 20 mv, 50 mv, 100 mv, 200 mv, 500 mv, 1 v ? 3 ? v ertical resolution for 8 bits = 0 ? 4 % of full scale, for 12 bits = 0 ? 0 24% of full scale ? 4 ? 1 3 ghz dsp enhanced bandwidth not applicable at 5 mv/div ? ) )7dpsolwxghuhdglqjvduhdiihfwhgevfrshdqgsurehedqgzlgwkolplwdwlrqvdqglqsxwdpsolohuvurooriihjg%urooriidwvshflohgedqgzlgwkri scope/probe) ? 6 ? t he fft signal to noise ratio varies with volts/division setting, memory depth and use of time or frequency averaging ? d 1 rlvhlvwkhglvsodhgqrlvhprru6ohz5dwhlvwkhglvsodhgvohzudwhriwkhvljqdodwwkhwkuhvkrogfurvvlqjv6dpsohudwh pdvlqlqwhusrodwlrq enabled ? e 0 hdvxuhphqwwkuhvkrog ohgyrowdjhdwohyho f 7 lphudqjhvv 8 ? v alues represent time error between two edges on a single channel ? standard deviation value refers to the standard deviation of 256 consecutive phdvxuhphqwvshuiruphgxvlqjdqlqglylgxdolqvwuxphqw5hdglqjlvwkhglvsodhg70phdvxuhphqwydoxh7lph6fdohfflvwkhrvfloorvfrshjvvshflohg time scale accuracy ? , qwhuqdohgjhwuljjhuprgh7uljjhuwkuhvkrog ohgyrowdjhdwohyho7khvohzudwhlqghshqghqwydoxhlqwkhirupxoduhsuhvhqwvwkhwudglwlrqdowuljjhu jit ter ? 5 htxluhv2swlrqxvhughoqhgixqfwlrq 11 ? 1 1 ? 8 g hz analog bandwidth at 5 mv/div for dso91304a and dso91204a models ? 12 ? t ypically triggers as low as 5 mv/div sensitivity ? 13 ? m easurement histograms require ezjit license ? infiniium 90000a series oscilloscopes (continued)
34 | keysight | infiniium 90000 series oscilloscopes - data sheet infiniimax ii series performance characteristics 1169a, 1168a bandwidth 1 1169a: > 12 ghz (13 ghz typical) 1168a: > 10 ghz rise and fall time probe only 1169a: 28 ps (20 to 80%), 40 ps (10 to 90%) 1168a: 34 ps (20 to 80%), 48 ps (10 to 90%) when phase compensated by 1169a with 91204a: 25 ps (20 to 80%) 1168a with 90804a: 38 ps (20 to 80%) 36 ps (10 to 90%) 54 ps (10 to 90%) 90000a series oscilloscope 1169a with 91304a: 23 ps (20 to 80%) 33 ps (10 to 90%) system bandwidth (C3 db) 1169a with 91304a: 13 ghz (typical) 1168a with 90804a: 8 ghz 1169a with 91204a: 12 ghz input capacitance 2 cm = 0.09 pf cm is between tips cg = 0.26 pf cg is to ground for each tip cdiff = 0.21 pf differential mode capacitance = cm + cg/2 cse = 0.35 pf single-ended mode capacitance = cm + cg input resistance 1 differential mode resistance = 50 k? 2% single-ended mode resistance = 25 k? 2% input dynamic range 3.3 v peak to peak, 1.65 v input common mode range 6.75 v peak to peak dc to 100 hz; 1.25 v peak to peak > 100 hz maximum signal slew rate 25 v/ns when probing a single-ended signal 40 v/ns when probing a differential signal dc attenuation 3.45:1 zero offset error referred to input 1.5 mv offset range 16.0 v when probing single-ended offset gain accuracy < 1% of setting when probing single-ended noise referred to input 2.5 mv rms, probe only propagation delay ~6 ns (this delay can be deskewed relative to other signals) maximum input voltage 30 v peak, cat i esd tolerance > 8 kv from 100 pf, 300 ? hbm temperature operating: 5 c to +40 c non-operating: 0 c to +70 c 1 ? denotes warranted specifications, all others are typical ? 2 ? measured using the probe amplifier and n5381a solder-in differential probe head ?
35 | keysight | infiniium 90000 series oscilloscopes - data sheet infiniimax ii series (continued) performance characteristics 1134a, 1132a, 1131a, 1130a bandwidth 1 1134a: > 7 ghz 1131a: > 3.5 ghz 1132a: > 5 ghz 1130a: > 1.5 ghz rise and fall time (10% to 90%) 1134a: 60 ps 1131a: 100 ps 1132a: 86 ps 1130a: 233 ps system bandwidth (C3 db) 1134a with 90604a: 6 ghz 1132a with 90404a: 4 ghz 1131a with 90254a: 2.5 ghz input capacitance 2 cm = 0.10 pf cm is between tips cg = 0.34 pf cg is to ground for each tip cdiff = 0.27 pf differential mode capacitance = cm + cg/2 cse = 0.44 pf single-ended mode capacitance = cm + cg input resistance 1 differential mode resistance = 50 k? 2% single-ended mode resistance = 25 k? 2% input dynamic range 5.0 v peak to peak, 2.5 v input common mode range 6.75 v peak to peak dc to 100 hz; 1.25 v peak to peak > 100 hz maximum signal slew rate 18 v/ns when probing a single-ended signal 30 v/ns when probing a differential signal dc attenuation 10:1 3% before calibration on oscilloscope 10:1 1% after calibration on oscilloscope zero offset error referred to input < 30 mv before calibration on oscilloscope < 5 mv after calibration on oscilloscope offset range 12.0 v when probing single-ended offset accuracy < 1% of setting when probing single-ended noise referred to input 3.0 mv rms propagation delay ~6 ns (this delay can be deskewed relative to other signals) maximum input voltage 30 v peak, cat i esd tolerance > 8 kv from 100 pf, 300 ? hbm temperature operating: 5 c to +40 c non-operating: 0 c to +70 c 1 ? denotes warranted specifications, all others are typical ? 2 ? measured using the probe amplifier and n5381a solder-in differential probe head ?
36 | keysight | infiniium 90000 series oscilloscopes - data sheet infiniium dsa/dso90000a series oscilloscopes model bandwidth channels sample rate standard memory per channel dsa/dso91304a 13 ghz 4 40 gsa/s 20 mpts/50 mpts (dsa) dsa/dso91204a 12 ghz 4 40 gsa/s 20 mpts/50 mpts (dsa) dsa/dso90804a 8 ghz 4 40 gsa/s 20 mpts/50 mpts (dsa) dsa/dso90604a 6 ghz 4 20 gsa/s 20 mpts/50 mpts (dsa) dsa/dso90404a 4 ghz 4 20 gsa/s 20 mpts/50 mpts (dsa) dsa/dso90254a 2.5 ghz 4 20 gsa/s 20 mpts/50 mpts (dsa) standard accessories usb optical mouse C high-performance calibration cable (not included in dsa/ dso90254a) usb keyboard C e2655b probe deskew and performance verification kit users quick-start guide C two 54855-67604 bnc-compatible to precision 3.5 mm (f) adapters (not included in dsa/dso90254a) power cord stylus pen dsa series comes with standard 50 mpts memory, high speed serial data analysis (option 003/e2688a), ezjit complete jitter analysis software (option 070/n8823a), ezjit plus jitter analysis software (option 004/n5400a), and ezjit jitter analysis software (option 002/e2681a) ? note: no probes are included with the dsa/dso90000a series oscilloscopes ? the infiniimax series probes or any other probes must be purchased separately ? additional options and accessories dso90000a-1cm rack mount kit dso90000a-a61 ansi z540 compliant calibration dso90000a-801 removable solid state drive n2892a (requires option 801) additional removable solid state drive for 90000 series with windows 7 dso90000a-805 gpib card-interface dso90000a-807 1 m?, adapter with a 500 mhz passive probe dso90000a-820 dvd-rw dso90000a-821 additional precision bnc to sma adapters, qty 2 for post sales, order two 54855-67604 precision bncs dso90000a-822 external touchscreen monitor for infiniium dso90000a-1a7 iso17025 compliant calibration presales memory options dso90000a-20m 20 mpts memory/ch upgrade dso90000a-50m 50 mpts memory/ch upgrade 1 dso90000a-100 100 mpts memory/ch upgrade dso90000a-200 200 mpts memory/ch upgrade dso90000a-500 500 mpts memory/ch upgrade dso90000a-01g 1 gpts memory/ch upgrade 1 ? standard on dsa version oscilloscopes ? mount your 90000 series scope in a 19 (487 mm) rack with option 1cm ? quickly remove your hard drive for additional security with option 801 ? infiniium 90000 series ordering information
37 | keysight | infiniium 90000 series oscilloscopes - data sheet factory installed option for new purchases user installed standalone product number sw applications 002 e2681a-1nl ezjit jitter analysis software (standard on dsa series) 003 e2688a-1nl high-speed serial data analysis with clock recovery and 8b/10b decoding (standard on dsa series) 004 n540 0a-1nl ezjit plus jitter analysis software (standard on dsa series) 007 n5391a-1nl protocol triggering and decode i 2 c/spi 009 n5414b-1nl infiniiscan event identification software 010 n54 30a-1nl infiniium user-defined function application software 012 n5461a-1nl serial data equalization 013 n5465a-3nl basic infiniisim waveform transformation toolset 014 n5465a-1nl advanced infiniisim waveform transformation toolset 015 n5462a-1nl rs-232/uart protocol triggering and decode 016 n5464a-1nl usb 2.0 protocol triggering and decode 017 n5463a-1nl pci express protocol triggering and decode 018 n8 801a-1nl sas/sata protocol triggering and decode 019 n8 802 a-1nl mipi d-phy protocol triggering and decode 021 n5392b-3nl ethernet electrical performance validation and compliance software 022 n5393f-1fp pci express electrical performance validations and compliance software 023 n5399c-3nl hdmi 1.4 electrical performance validation and compliance software 029 n5416a-1nl usb 2.0 compliance test software 030 n5431a-1nl xaui electrical validation with 10gbase-cx4, cpri, obsai, and serial rapidio 031 u7233a-1nl ddr1 and lpddr compliance test applications 032 n5413b-1nl ddr2 and lpddr2 compliance test applications 033 u7231b-1nl ddr3 and lpddr3 compliance test applications 034 n5394a dvi compliance application 035 u7238a mipi compliance test application 036 u7236a 10gbase-t ethernet electrical compliance application 038 n5411b-1nl sata 6g compliance test software 040 n5467b-1nl user definable application 041 u7243b-3nl usb 3.1 compliance test software 043 N5412D-3NL serial attached scsi (sas-2) electrical performance validation and compliance 045 u7232c-1nl displayport 1.2 compliance test software 058 n6 462 a-1nl ddr4 and lpddr4 compliance test application 060 n5392b 10/100/1000bte energy efficient ethernet 061 matlab - basic digital analysis package 062 matlab - standard digital analysis package 063 n8803a-1nl can, lin and flexray protocal triggering and decode 065 n6 467a-1nl broadr-reach compliance 070 n8 823a-1nl ezjit complete jitter and noise analysis software (standard on dsa series) 073 n6466a-1nl most compliance n8 8 45a-1fp n8845a-2fp spmi protocol trigger and decode 1 1 generic raw C nrz protocol decode 2 2 generic raw C pam-4 protocol decode 1 ? standard with sda option (firmware 6 ? 0 a nd up) ? 2 ? requires pam-4 compliance app and sda option (firmware 6 ? 0 a nd up) ? infiniium 90000 series ordering information (continued)
38 | keysight | infiniium 90000 series oscilloscopes - data sheet oscilloscope bandwidth upgrades upgrade descriptions n5471a dsa/dso91204a to dsa/dso91304a upgrade (12 to 13 ghz) n5471b dsa/dso90804a to dsa/dso91204a upgrade (8 to 12 ghz) n5471c dsa/dso90604a to dsa/dso90804a upgrade (6 to 8 ghz) n5471d dsa/dso90404a to dsa/dso90604a upgrade (4 to 6 ghz) n5471e dsa/dso90254a to dsa/dso90404a upgrade (2.5 to 4 ghz) note: order as many upgrades as needed to reach the desired final bandwidth of the instrument ? for example, to upgrade from a dsa/dso90804a to dsa/dso91304a order n5471b and n5471a ? oscilloscope memory upgrades upgrade descriptions n5472a after-purchase 10 mpts per channel to 20 mpts per channel memory upgrade n5472b after-purchase 20 mpts per channel to 50 mpts per channel memory upgrade n5472c after-purchase 50 mpts per channel to 100 mpts per channel memory upgrade n5472d after-purchase 100 mpts per channel to 200 mpts per channel memory upgrade n5472e after-purchase 200 mpts per channel to 500 mpts per channel memory upgrade n5472f after-purchase 500 mpts per channel to 1 gpts per channel memory upgrade operating system upgrades upgrade descriptions n2753a windows 7 for infiniium 90000 scope with windows xp and sn>my50410100 n2754a-002 window 7 and m890 motherboard for infiniium 90000 scopes with windows xp and sn 39 | keysight | infiniium 90000 series oscilloscopes - data sheet keysight oscilloscopes multiple form factors from 20 mhz to > 90 ghz | industry leading specs | powerful applications
40 | keysight | infiniium 90000 series oscilloscopes - data sheet www.axiestandard.org advancedtca ? extensions for instrumentation and test (axie) is an open standard that extends the advancedtca for general purpose and semiconductor test ? the business that became keysight was a founding member of the axie consortium ? atca ? , advancedtca ? , and the atca logo are registered us trademarks of the pci industrial computer manufacturers group ? www.lxistandard.org lan extensions for instruments puts the power of ethernet and the web inside your test systems ? the business that became keysight was a founding member of the lxi consortium ? www.pxisa.org pci extensions for instrumentation (pxi) modular instrumentation delivers a rugged, pc-based high-performance measurement and automation system ? learn more at www.keysight.com/find/software start with a 30-day free trial. www.keysight.com/find/free_trials download your next insight keysight software is downloadable expertise ? from first simulation through first customer shipment, we deliver the tools your team needs to accelerate from data to information to actionable insight ? C electronic design automation (eda) software C application software C programming environments C productivity software
41 | keysight | infiniium 90000 series oscilloscopes - data sheet this information is subject to change without notice ? ? keysight technologies, 2012 - 2017 published in usa, september 13, 2017 5989-7819en www.keysight.com evolving since 1939 our unique combination of hardware, software, services, and people can help you reach your next breakthrough ? we are unlocking the future of technology ? from hewlett-packard to agilent to keysight ? mykeysight www.keysight.com/find/mykeysight a personalized view into the information most relevant to you ? http://www.keysight.com/find/emt_product_registration register your products to get up-to-date product information and find warranty information ? keysight services www.keysight.com/find/service keysight services can help from acquisition to renewal across your instruments lifecycle ? our comprehensive service offeringsone- stop calibration, repair, asset management, technology refresh, consulting, training and morehelps you improve product quality and lower costs ? keysight assurance plans www.keysight.com/find/assuranceplans up to ten years of protection and no budgetary surprises to ensure your instruments are operating to specification, so you can rely on accurate measurements ? keysight channel partners www.keysight.com/find/channelpartners get the best of both worlds: keysights measurement expertise and product breadth, combined with channel partner convenience ? cdma2000 is a us registered certification mark of the telecommunica- tions industry association ? pci-sig ? , pcie ? and the pci express ? are us registered trademarks and/or service marks of pci-sig ? mipi ? service marks and logo marks are owned by mipi alliance, inc ? and any use of such marks by keysight technologies is under license ? other service marks and trade names are those of their respective owners ? wimax, mobile wimax, wimax forum, the wimax forum logo, wimax forum certified, and the wimax forum certified logo are us trademarks of the wimax forum ? for more information on keysight technologies products, applications or services, please contact your local keysight office ? the complete list is available at: www.keysight.com/find/contactus americas canada (877) 894 4414 brazil 55 11 3351 7010 mexico 001 800 254 2440 united states (800) 829 4444 asia pacific australia 1 800 629 485 china 800 810 0189 hong kong 800 938 693 india 1 800 11 2626 japan 0120 (421) 345 korea 080 769 0800 malaysia 1 800 888 848 singapore 1 800 375 8100 taiwan 0800 047 866 other ap countries (65) 6375 8100 europe & middle east austria 0800 001122 belgium 0800 58580 finland 0800 523252 france 0805 980333 germany 0800 6270999 ireland 1800 832700 israel 1 809 343051 italy 800 599100 luxembourg +32 800 58580 netherlands 0800 0233200 russia 8800 5009286 spain 800 000154 sweden 0200 882255 switzerland 0800 805353 opt ? 1 (de) opt ? 2 (fr) opt ? 3 (it) united kingdom 0800 0260637 for other unlisted countries: www.keysight.com/find/contactus (bp-9-7-17) dekra certified iso9001 quality management system www.keysight.com/go/quality keysight technologies, inc ? dekra certified iso 9001:2015 quality management system


▲Up To Search▲   

 
Price & Availability of N5412D-3NL

All Rights Reserved © IC-ON-LINE 2003 - 2022  

[Add Bookmark] [Contact Us] [Link exchange] [Privacy policy]
Mirror Sites :  [www.datasheet.hk]   [www.maxim4u.com]  [www.ic-on-line.cn] [www.ic-on-line.com] [www.ic-on-line.net] [www.alldatasheet.com.cn] [www.gdcy.com]  [www.gdcy.net]


 . . . . .
  We use cookies to deliver the best possible web experience and assist with our advertising efforts. By continuing to use this site, you consent to the use of cookies. For more information on cookies, please take a look at our Privacy Policy. X